Radiation Environment and Medicine
Online ISSN : 2432-163X
Print ISSN : 2423-9097
ISSN-L : 2423-9097
Note
Low-volume Electrolytic Enrichment for Tritium Measurement Using Improved Solid Polymer Electrolyte System at NIFS and Its Application
Naofumi AkataChie IwataAkemi KatoMasahiro TanakaHirofumi TazoeNagayoshi ShimaKimpei IchiyanagiMiklós HegedűsGergő BátorTibor KovácsHideki Kakiuchi
Author information
JOURNAL FREE ACCESS

2020 Volume 9 Issue 2 Pages 93-97

Details
Abstract

We evaluated the enrichment factor of tritium for an improved solid polymer electrolyte (SPE) system. In this system, water sample reservoirs were made of double-glazed glass, and cooling water was circulated in the double glass to cool the sample. As the result, a tritium enrichment factor of 5.00 was obtained and the MDL of tritium measurement with a low-background liquid scintillation counter coupled to this improved enrichment process was determined to be approximately 0.062 Bq L-1. Tritium concentration in monthly precipitation at Kumamoto was measured using this method combining a low-background liquid scintillation counter and this improved SPE system.

Content from these authors
© 2020 Hirosaki University Press.
Previous article Next article
feedback
Top